  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 49
50
4
CALIBRATION
*
Applicable to a veried balance as a legal measuring instrument.
Calibration with internal weight executes automatically in weighing mode under any of the
following circumstances.
(1) When there is a change in the surrounding temperature.
(2) When about four hours has passed since the previous calibration.
(3) After either condition (1) or (2) above has been met during warm up at stand-by status, when the
balance is switched to mass display mode.
When there is a change in the surrounding
temperature or about four hours has passed
since the previous span adjustment, the
indicator will blink.
When two minutes pass since blinking
start,
Case1) There is no load on the pan.
"PSC.RUN" is displayed and automatic span
adjustment is executed.
It returns to the weighing mode when automatic
span adjustment ends.
Case 2) There is some load on the pan.
Mass display will blinks.
When two minutes pass since mass display
blinking start, "PLS.CAL" is displayed.
If "PLS.CAL" is displayed, please unload the
thing on the pan and execute span adjustment.
*
Push key when you must keep measuring after “PLS.CAL” is displayed. Return to
blinking mass display.
If the span adjustment is not executed, the mass display blinking for two minutes and "PLS.CAL"
display are repeated.
! !0.0!0!0!0
!P!S!C.R!U!N
! !0.0!0!0!0
!P!L!S.C!A!L
! !0.0!0!0!0
!P!S!C.R!U!N
! !0.0!0!0!0
!P!L!S.C!A!L
Automatic
calibration
(ATX only)
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/