5
FUNCTIONS RELATING TO TARING
When the auto zero function is set, any drift of the zero point that occurs as a result of material left on
the pan after weighing is automatically compensated for so that zero is displayed.
Note that the auto zero function cannot be used in combination with formulation.
*
Not applicable to a veried balance as a legal measuring instrument.
1
Press for about
3 seconds in the weighing mode.
This opens the zero / taring menu.
2
Select the auto zero function.
[A.ZERO]
What is the current situation? What do you want to do?
Stability
Mark
Auto Zero
Function
To set or Update… To Cancel…
Lit
ID is ON
Press and go to step 3.
Press and go to
step 4.
Unlit
ID is OFF
Press and go to
step 3.
Go to step 4.
3
Enter the value for the range for
automatic compensation to the
zero point (auto zero range).
(Enter the zero range value.) [SET]
^
"Entering Numerical Values", page 43
If there is anything with a weight lower than or
equal to the auto zero range value left on the pan
after weighing the sample, it will automatically
be compensated for and the zero point will be
established when (the stability mark) lights
up.
! !A.Z!E!R!O
Check the presence or absence of
the stability mark.
! !A.Z!E!R!O
Check the presence or absence of
the stability mark.
! !0.0!0!0!0
! !S!E!T! !
! !A.Z!E!R!O
/
/
/
/
/
/
/
/
/
/
/
/
/
/
When ON is set the stability mark is lit.
! !0.0!0!0!0
! !S!E!T! !
! !A.Z!E!R!O
/
/
/
/
/
/
/
/
/
/
/
/
/
/
When ON is set the stability mark is lit.