5.2 Setting the Auxiliary Functions
5-19
RF-20A/20Axs
5
Setting the Sensitivity [SENS]
Set the sensitivity of the instrument.
^ "4.1.8 Setting Sensitivity" P.4-17
"4.2.8 Setting Sensitivity" P.4-38
Enter the set value from the numeric keys and press
.
Setting the Gain [GAIN]
Set a magnification of the sensitivity (SENS). By
combining the setting of sensitivity and gain, you can set
the measuring range.
^ "4.1.7 Setting Gain" P.4-16
"4.2.7 Setting Gain" P.4-36
Enter the set value from the numeric keys and press
.
The combinations of sensitivity and gain are shown
below.
NOTE
When the sensitivity and/or gain have been changed,
plot a revised calibration curve.
SENS 2
1:HI 2:MED 3:LOW
Set Value Sensitivity Sensitivity Magnifications
1 HI Approx. × 1
2 MED Approx. × 32
3 LOW Approx. × 1024
GAIN 2
1:x1 2:x4 3:x16
Set Value Gain
1 × 1
2 × 4
3 × 16
Sensitivity Gain Sensitivity Magnifications
3 (LOW)
1 Approx. × 1
2 Approx. × 4
3 Approx. × 16
2 (MED)
1 Approx. × 32
2 Approx. × 128
3 Approx. × 512
1 (HI)
1 Approx. × 1024
2 Approx. × 4096
3 Approx. × 16384
Page view 118