8. Maintenance
8-18
RF-20A/20Axs
Fitting the Cell
NOTE
Fit the cell so that the inlet side and outlet side face
toward the front face of the flow cell unit.
Fitting the cell the wrong way round will cause liquid
leakage.
When the flow cell unit has been disassembled and
cleaned, be sure to replace the cell gasket.
Remove dust and so on from the new cell gasket
before using it.
Fig. 8.24
RF-20A
1
Fit the cell in the cell housing by following the
procedure for removal in reverse.
Secure the cell gasket, cell and spacer while they
are pressed against the lower part of the cell
housing.
Tighten the two screws of the rear plate,
alternately by equal amounts, half a turn at a time.
Fig. 8.25
RF-20Axs
Fig. 8.26
Cell housing
Cell gasket
Cell
Fit the cell with the inlet/outlet facing the
front face of the cell housing.
The illustration shows the RF-20Axs.
Inlet/outlet
Cell housing
Screw
Rear plate
Spacer
Cell
Cell gasket
Cell housing
Cell gasket
Cell
Spacer
Rear plate
Screw