5. Application Operation
5-64
RF-20A/20Axs
5.5.4 Setting the Start Wavelength and End Wavelength
Set the start wavelength and end wavelength for spectrum scanning.
If [1] (excitation scanning) was set in "5.5.3 Setting the Scan Type" P.5-63 the excitation wavelength is set,
and if [2] (emission scanning) was set the emission wavelength is set. If [1] (excitation scanning) was set, [EX
SCAN] is displayed. If [2] (emission scanning) was set, [EM SCAN] is displayed. The setting ranges for each
wavelength are as follows.
Excitation wavelength setting screen
1
Press repeatedly until [EX SCAN] or [EM
SCAN] in the control settings group is displayed.
If [1] was set for the scan type display [EX
SCAN], and if [2] was set for the scan type
display [EM SCAN].
^ "5.5.3 Setting the Scan Type" P.5-63
Emission wavelength setting screen
2
In the initial status, the start wavelength can be
entered.
Enter the start wavelength with the numeric keys
and press .
It will now become possible to enter the end
wavelength.
3
Enter the end wavelength with the numeric keys
and press .
Wavelength
Setting Range (Units: nm)
Start Wavelength End Wavelength
EX (excitation
wavelength)
200 to 900 200 to 900
EM (emission
wavelength)
200 to 900 200 to 900
EX SCAN 200-900
Input 200 - 900
EM SCAN 200-900
Input 200 - 900