7.5 Detector Validation
7-15
RF-20A/20Axs
7
When a system controller or LCsolution is used, perform the following operation before analyzing the sample.
When the system controller is connected:
1
Turn the power to the system controller OFF.
2
Set [0] for [LOCAL] in the system settings group.
3
Restart the system controller.
When LCsolution is used:
1
Exit LCsolution.
2
Turn the power to the system controller OFF.
3
Set [0] for [LOCAL] in the system settings group.
4
After restarting the system controller, start LCsolution.
NOTE
Before starting analysis when using a system controller or LCsolution, be sure to set [0] for [LOCAL] in the
system settings group, then restart the system controller and LCsolution.