5.5 Measuring in the Spectrum Scanning Mode
5-67
RF-20A/20Axs
5
5.5.7 Setting Sensitivity
This is the procedure for setting sensitivity. By combining the setting of sensitivity and gain, you can set the
measuring range.
^ "4.1.8 Setting Sensitivity" P.4-17
The settings for sensitivity are shown below. On shipping from the factory, the setting is [2] (MED).
1
Press .
The initial screen will be displayed.
2
Press .
[PARAMETER] will be displayed.
3
Press .
[ch1] will be displayed.
4
Press repeatedly until [SENS] is
displayed.
5
Enter the set value from the numeric keys and
press .
This sets the sensitivity.
Set Value Sensitivity Sensitivity Magnifications
1 HIGH Approx. × 1
2 MED Approx. × 32
3 LOW Approx. × 1024
EX200nm EM300nm
Xe 1000.00
PARAMETER
Enter to Select
ch1 EX350 EM450
Input 0,200-900
SENS 2
1:HI 2:MED 3:LOW