6. Troubleshooting
6-6
RF-20A/20Axs
ERR EX HOME POS
Fatal
Cause: Displayed when the home position sensor of
the excitation grating motor is not operating
correctly.
(Home position sensor error)
Action: Turn the power OFF and then restart the
instrument. If the same error is still displayed,
contact your Shimadzu representative.
ERR EM HOME POS
Fatal
Cause: Displayed when the home position sensor of
the emission grating motor is not operating
correctly.
(Home position sensor error)
Action: Turn the power OFF and then restart the
instrument. If the same error is still displayed,
contact your Shimadzu representative.
ERR EEPROM WRITE
Fatal
Cause: An error in writing to the EEPROM has
occurred.
(EEPROM writing error)
Action: Turn the power OFF and then restart the
instrument. If the same error is still displayed,
contact your Shimadzu representative.
SYSTEM ERROR
Fatal
Cause: Displayed when an unanticipated error has
occurred in the internal system.
(SYSTEM error)
Action: Turn the power OFF and then restart the
instrument. If the same error is still displayed,
contact your Shimadzu representative.
ERR COOLER
Fatal
Cause: Displayed when the temperature of the flow cell
has failed to reach the set temperature within
30 minutes after the power was turned ON.
(Cooler error)
Action: Turn the power OFF and then restart the
instrument. If the same error is still displayed,
contact your Shimadzu representative.
Error Message Type Cause and Action