8.5 Replacing the Fuse
8-35
RF-20A/20Axs
8
8.5 Replacing the Fuse
Be sure to use fuses of the following type and capacity for replacement.
Parts used
8.5.1 Replacing the Fuse
1
Using a screwdriver, turn the fuse holder
counterclockwise while applying a little pressure
to remove it.
Fig. 8.51
2
Take the fuse out of the fuse holder.
Fig. 8.52
!WARNING
Before replacing fuses, turn the power to the instrument OFF and unplug the instrument.
Only use fuses of the correct type and rating for replacement.
Failure to heed the above could result in fire, electric shock or short circuits.
Part Name Part Type Part No.
250 V 5 AT (5 × 20) Replacement part 072-02004-23
Fuse holders
Screwdriver
Fuse
holder
Fuse
Page view 284