Shimadzu BL3200S Instruction Manual Page 210

  • Download
  • Add to my manuals
  • Print
  • Page
    / 352
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 209
7. Hardware Validation
7-12
RF-20A/20Axs
Check Procedure (for RF-20Axs)
NOTE
To discontinue the inspection, press .
1
When a system controller is being used, set [1] for
[LOCAL] in the system settings group.
2
Using the pump, pump water into the flow cell
(flow rate: 1 mL/min).
3
After checking that there are no air bubbles inside
it, fit the flow cell.
4
With the initial screen displayed, repeatedly press
until [VALIDATION] appears.
5
Press .
[DATE] will be displayed.
[EX CHECK]
6
Repeatedly press until [EX CHECK] or
[EM CHECK] appears.
[EM CHECK]
NOTE
For a wavelength accuracy check at the excitation side,
execute [EX CHECK], and for a wavelength accuracy
check at the emission side, execute [EM CHECK].
VALIDATION
Press func or VP
DATE YY-MM-DD
00-00-00
EX CHECK
Enter to Start
EM CHECK
Enter to Start