8. Maintenance
8-8
RF-20A/20Axs
8.2.2 Simple Cleaning of the Cell
If any air bubble or dirt is found on inspecting the cell, perform simple cleaning of the cell by following the
procedure below.
Parts used
Fig. 8.7
Part Name Part Type Part No.
Syringe Standard accessory 046-00017-01
Syringe adapter Standard accessory 228-15672-91
The arrows indicate the direction of flow of the liquid.
Cell outlet tube
Cell inlet tube
Blue tubing
Syringe adapter
Waste
container
Flow cell unit
Syringe