8.2 Inspection and Simple Washing of the Cell
8-9
RF-20A/20Axs
8
Simple Cleaning of the Cell
1
Fit the syringe adapter to the tip of the syringe.
Fig. 8.8
2
Remove the male nut PEEK at the column side of
the inlet coupling 1.6-0.8C, and remove the tubing
from the column side.
Fig. 8.9
3
Fit the male nut 1.6MN of the syringe adapter to
the inlet coupling 1.6-0.8C.
Fig. 8.10
Syringe
Syringe adapter
Blue
tubing
Inlet coupling
1.6-0.8C
Male nut, PEEK
To the waste
container
From the
column
Syringe
Inlet coupling
1.6-0.8C
Blue tubing
Male nut 1.6MN
Syringe adapter