5. Application Operation
5-82
RF-20A/20Axs
5.8.2 Wiring
1
Strip off 10 mm of insulation from the end of the
cable to be connected. Note that this is not
necessary with the event cable provided.
Fig. 5.4
2
If the cable comprises single core wires, insert the
wires into the holes in the terminals as they are.
If the cable comprises stranded wires, twist the
strands at the end of the wire together well, then
insert the wire with a tool such as a flat bladed
screwdriver while holding down the button at the
left of the terminal.
Also hold this button down when removing the
cable.
Signal Explanation Remark
EVENT1 (output) This is a relay contact output. It is set ON and OFF by the
time program or the EVENT setting in the auxiliary
functions.
The rating for the relay contact is
30 V DC / 1 A
EVENT2 (output)
RSVD Not used. Do not connect this terminal.
PROG. START
(input)
Starts this instrument's time program in response to an
external contact signal. If the start signal is input while the
time program is running, it is restarted from time "0".
These signals are controlled by
shorting between the input command
terminal and the COMMON terminal in
response to contacts at external
devices.
The during of shorting (tc) should be
as follows.
0.5 sec < tc < 10 sec
PROG. STOP
(input)
Stops this instrument's time program in response to an
external contact signal.
COMMON This is the COMMON terminal for inputs.
Strip off 10 mm of insulation.